Display the sensor ranges used by HART transmitter.
Address: F: MVAL: SENSOR_RANGE: Lower limit of sensor: Higher limit of sensor: Pressure unit
Switch to function 1,it is returned but not enter into the interface yet (example 1); oppositely, it returns after interface
displayed (example 2)
Example 1: 001: F: MVAL: SENSOR_RANGE: 0.00000000: 0.00000000: NC
Example 2: 001: F: MVAL: SENSOR_RANGE: -40.0000000: 40.0000000: KPA
Display and transfer the PV range of transmitter, switch PV units
Address: F: MVAL: PV_RANGE: Lower limit of PV range: Higher limit of PV range: Pressure unit
Switch to function 2, it returned but not enter in the PV interface yet (example 1); oppositely, it returns after interface
displayed (example 2)
Example------001: F: MVAL: PV_RANGE: 0.00000000: 0.00000000: NC
Example------001: F: MVAL: PV_RANGE: -0.01506382: -0.02472788: KPA
Display and adjust transmitter's damp value
Address: F: MVAL: DAMPING: damp value: Unit (second)
Example------001: F: MVAL: DAMPING: 1.000: s
Simultaneously display the pressure value measured by calibrator and transmitter.
Address: F: MVAL: PV/MP_VALUE: pressure measured by transmitter:Unit:pressure measured by calibrator:Unit
Switch to function 4, it returned but not enter in the PV/MP interface (example 1); oppositely, it returns after interface
displayed (example 2)
Example 1------001: F: MVAL: PV/MP_VALUE: 0.00000: NC: 0.01414: KPA
Example 2------001: F: MVAL: PV/MP_VALUE: 0.00973: KPA: 0.00984: KPA
Function 1:
Function 2:
Function 3:
Function 4: