62
(3) The file's transfer format
(4) Pressure units shortening list
Shortening
Standard
H O
2
mmH O
2
HG
mmHg
PSI
psi
MBAR
mbar
BAR
bar
PA
Pa
KPA
kPa
MPA
MPa
H O
2
HG
PSI
MBAR
BAR
PA
KPA
MPA
MSB LSB
(5) Pressure units info code
(6) The format of automatically data sending
use 1 byte.1: this pressure unit is available; 0: N/A
The format length is 32 bytes, and add the Eos 0x0 at the end.
Info code
Filename: F01 /* File name */
Number: 111111 /* the serial No. of second gauge */
Minscale: 001.000
/* Minimum scale value */
Datesum: 01
/* the saved points in file */
No.01 06/07/03 10:29:53
/* Point number, save date, time */
0.0108MPA /* Pressure measure value */
-0.0000mA
/* Electricity measure value */
Example: *P 0.0364 MPA *I -0.0001 mA
*P 0.0367 MPA *V -0.0158 V
*P 0.0374 MPA *T 32.19
℃
*P 0.0375 MPA *S 000000.0 0
*P 0.0397 MPA *L 10:00:05