Figure 5-3-2: 1 point calibration Figure 5-3-3: 2 point entrance Figure 5-3-4: 2 calibration
st
nd
nd
0
25
50
75
100
standard
pressure
calibration point
measuring
pressure
0
25
50
75
100
calibration point
standard pressure
0
25
50
75
100
calibration point
standard
pressure
measuring
pressure
(3) The first line flashes to remind the user to input the standard pressure of 1 point. (In this example, it is 0.00000 MPa)
(4) As figure 5-3-2, screen displays the interface of 1 calibration point. Please supply 0.00000 MPa pressure, and press
to finish the 1 point calibration after it is stable. As figure 5-3-3, switch to the 2 point calibration interface.
(5) Under the interface of 2 calibration point, the first line flashes for new standard pressure. In general, it is the default (In this
example, it is 2.00000 MPa), press for next step.
(6) As figure 5-3-4, screen displays the 2 point calibration interface. Please supply 2.00000 MPa pressure and press to
finish the 2 point calibration after it is stable. The screen switches to the interface as figure 5-3-1 and the calibration is
finished.
(7) Press to return to the calibration function menu. The calibration is effective if changed to .
st
st
st
nd
nd
nd
nd
51