21
3.1
TC Measurement
There are 2 mini-TC sockets on the top of the unit, corresponding to two different channels. After connecting the
TC to the device, click the type switch menu
or
of the corresponding temperature channel and select the TC to
execute the TC measurement function.
3.2
TC Measurement Setup
Click “Setup” in the TC measurement menu to enter the TC setup. The following parameters can be set in this menu.
Table 7 TC Measurement Setup
Menu Selection
Effective value
Description
TC types
mV/TC sensors
Choose the TC sensors
Cold junction type
Internal/External/Custom
Choose the mode of cold junction. Built-in cold
junction of device is used for Internal mode.
External cold junction compensation range
-
80~300
℃
Apply to external cold junction mode.
Temperature unit
K/
℉
/
℃
Choose temperature unit
Resolution
0/0.1/0.01/0.001
Number of displayed decimal places
3.3