12707 Resistivity Chamber for Liquid Sample Operation Manual
2
.2
Part Descriptions
z
See Figure 2-3.
(10)
Ring electrode
A container to put a measurement sample in. Voltage is applied to this electrode.
(11)
Internal electrode
This electrode is used in the ring electrode. It contains the main electrode and the guard
electrodes.
(12)
Main electrode
An electrode for measurement. It is connected with the M terminal (15).
(13)
Guard electrode A
This electrode cancels current that flows around from the top of the main electrode. Con-
nects with the G terminal (16).
(14)
Guard electrode B
This electrode cancels current that flows around from the bottom of the main electrode.
Connects with the G terminal (16).
(15)
M terminal
It is connected with the main electrode (12) for measurement. Connects with the MEAS-
URE terminal (5) by using the supplied cable (DCB-PP1801X01).
(16)
G terminal
It is connected with the guard electrode A (13) and the guard electrode B (14). Connects
with the GUARD terminal (6) by using the supplied cable (DCB-PP1801X01).
(17)
V terminal
A terminal of the ring electrode to apply voltage. Connects with the V SOURCE terminal
(7) by using the supplied cable (DCB-PP1801X01).