2-5
12707 Resistivity Chamber for Liquid Sample Operation Manual
2
.2
Part Descriptions
z
See Figure 2-2.
(4)
Electrode insertion opening
The electrode is inserted into this opening. The projections in the inside are aligned with
the holes on the bottom of the electrode.
(5)
MEASURE terminal
A terminal for measurement that connects with the M terminal (12) by using the supplied
cable (DCB-PP1801X01).
The insulation degradation of the MEASURE terminal has an influence directly on
measured values, causing measurement errors. If the terminal area gets dirty, clean it
with a swab soaked in absolute alcohol.
(6)
GUARD terminal
A terminal to measure precisely the volume resistivity that connects with the G terminal
(13) by using the supplied cable (DCB-PP1801X01).
(7)
V SOURCE terminal
A terminal to apply voltage that connects with the V terminal (17) by using the supplied
cable (DCB-PP1801X01).
(8)
LID SIGNAL switch
Turns on when the cover is closed.
CAUTION