Adam WBW 16 L User Manual Download Page 16

© Adam Equipment Company 2018

 

 

15

 

14.0  ERROR CODES 

 
During  the  initial  power-on  testing  or  during  operation,  the  scale  may  show  an  error 
message.  The meaning of the error messages is described below. 
 
If an error message is shown, repeat the step that caused the message. If the error message 
is still shown then contact your dealer for support. 
  
 

ERROR 
CODE 

DESCRIPTION 

POSSIBLE CAUSES 
 

Err  4 

Initial Zero is greater than 
allowed (4% of maximum 
capacity) when power is 
turned on or when 

[Zero]

 is 

pressed. 

Weight on the pan when turning the scale 
on. 
Excessive weight on the pan when zeroing 
the scale. 
Platform is not installed. 
Improper calibration of the scale. 
Damaged load cell. 
Damaged Electronics. 

Err 6 

A/D count is not correct 
when turning the scale on. 

Load cell is damaged. 
Electronics is damaged. 

Err  8 

High limit input error 

Low limit is set first, then the high limit is set 
lower than the low limit and high limit not 
equal to zero. 

Err 9 

Low limit input error 

High limit is set first, then the low limit is set 
higher than the high limit and low limit not 
equal to zero. 

FAIL H 

or

  

FAIL L 

Calibration error 

Improper calibration (should be 10% 
of the factory calibration). The old calibration 
data will be retained until the calibration 
process is complete. 

 

 

 

Summary of Contents for WBW 16 L

Page 1: ...Adam Equipment Company 2018 Adam Equipment WBW L Software revision U 4 01...

Page 2: ...SYMBOLS AND INDICATORS 8 7 0 BATTERY OPERATION 8 8 0 DISPLAY BRIGHTNESS 9 9 0 AUTO POWER OFF 9 10 0 OPERATIONS 10 10 1 ZEROING 10 10 2 TARING 10 10 3 WEIGHING 10 10 4 CHECK WEIGHING 11 10 5 LIMITS STO...

Page 3: ...steel weighing platforms on an ABS plastic base assembly which is sealed to IP 65 making it water proof All scales have sealed keypads with colour coded membrane switches and the displays are clear re...

Page 4: ...ity S D 0 2 g 0 5 g 1 g 2 g Linearity 0 4 g 1 g 2 g 4 g Pounds Maximum Capacity 5 lb 9 lb 18 lb 35 lb Tare Range 5 lb 9 lb 18 lb 35 lb Readability 0 0005 lb 0 001 lb 0 002 lb 0 005 lb Repeatability S...

Page 5: ...ugh an external Power Supply Module Battery Internal rechargeable battery 50 hours operation Calibration Automatic External Display Digit LED display with indication of functions Scale Housing IP 65 S...

Page 6: ...duce the accuracy Avoid extremes of temperature Do not place in direct sunlight or near air conditioning vents Avoid unsuitable tables The table or floor must be rigid and not vibrate Avoid unstable p...

Page 7: ...le in the spirit level is in the centre of the level and the scale is supported by all four feet Attach the power supply module to the bottom of the scale and plug into the mains Press the O I key to...

Page 8: ...and shows the results This is the net weight Enters the selected parameter or value for setting Unit Selects the weighing units from a preset list of available units Shifts the flashing digit to the n...

Page 9: ...isplay to the left there are three LED s that indicates when the weight is below within or over the pre set limits during check weighing Weight LED Below the low limit Red Within the limits Green Abov...

Page 10: ...r the internal battery will be charged If the LED is green the battery is being charged If it is red it is nearly discharged and yellow indicates the battery is increasing the charge level Continue to...

Page 11: ...The stable and Net indicator will be on As a product is added only the weight of the product will be shown The scale could be tared a second time if another type of product was to be added to the fir...

Page 12: ...other than 0000 See the parameter F4PS in section 12 1 Enter the correct password using the numeric entry method as mentioned in section 4 1 If the password is 0000 it will display the last used high...

Page 13: ...weighing unit selected is either pounds ounces or pounds ounces Method To start the calibration turn the scale off and then turn it on again Press Tare during the self test Scale will show P Enter cod...

Page 14: ...ned in section 4 1 If the password is 0000 it will go directly to F1 LLK Parameter Description Options Default setting F1 LLt Limit Lock will prevent a normal user from changing the check weighing lim...

Page 15: ...e or disable weighing units will not allow to disable all units at least one has to be enabled Press enter to scroll through units to set kg g lb oz lb oz kg S2 bL Display brightness settings Br 1 Br...

Page 16: ...e scale on Excessive weight on the pan when zeroing the scale Platform is not installed Improper calibration of the scale Damaged load cell Damaged Electronics Err 6 A D count is not correct when turn...

Page 17: ...eady A Details of your company Name of your company Contact person s name Contact telephone e mail fax or any other methods B Details of the unit purchased This part of information should always be av...

Page 18: ...nal packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performanc...

Page 19: ...ercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio...

Page 20: ...49 0 4340 40300 0 Fax 49 0 4340 40300 20 e mail vertrieb aeadam de Adam Equipment S A Pty Ltd 7 Megawatt Road Spartan EXT 22 Kempton Park Johannesburg Republic of South Africa Phone 27 0 11 974 9745 F...

Reviews: