© Adam Equipment Company 2018
12
10.5 LIMITS STORED IN MEMORY
If the scale is turned off it stores the last high and low limits in the memory along with
information about the weighing unit in use when the limits were stored. If the scale is turned
on again, the limits and the weighing unit will be active.
11.0 CALIBRATION
The
WBW L
scales are calibrated using metric weights when the weighing unit selected is
either kilograms or grams and using pound masses when the weighing unit selected is either
pounds, ounces or pounds:ounces.
Method
To start the calibration turn the scale off and then turn it on again. Press
[Tare]
during the self-test. Scale will show “P- - - - “. Enter code number “
0000
” using
the numeric entry method (see section 4.1) and press
[Tare].
This will take
you directly to the calibration section.
Display will show "
UnLoAd
".
Press the
[Tare]
key; display will show “
Ld xx’’
(xx- weight value). Add the
suggested weight to the pan.
Press the
[Tare]
key and wait until the stability light is shown; then press
[Tare]
key again.
Display will show ‘’Pass’’ or ‘’Fail’’, then return to the weighing mode.
Remove the calibration weight when the calibration is complete.