Once the VI has started, the pulse waveform will be uploaded into the FPGA look-up tables on all
available channels
Press the
RUN
button to start the waveform generation
DDS and Trigger Route.vi
Software requirements: LabView 2012
This VI gives the user access to the clock generated by PXIe-6674T NI Timing board and to the
DSTARA/DSTARB clock/global trigger routing.
Set
the DDS frequency
to 156.25 MHz and fill the
Destination Terminal Array.
The Destination
Terminal Array should contain all the PXIe_DSTARA connection to the FlexRIO boards that
will provide the same reference clock to the adapter module (refer to the chassis manual).
Fill the
Destination Trigger Array
with the PXIe_DSTARB global trigger routing connections.
The FlexRIO boards will receive now the same trigger signal.
Press the
Send Trigger
button to send it to all connected FAMs.