33
Type
Description
Biolight Model
M70C
Dimensions of main unit
57mm (Length) x 33mm (Width) x 30mm (Height)
Weight
54g (including batteries)
Communication interface
Bluetooth Smart (4.0)
Sensor Specifications
Wavelength
Red light ~660nm
Infrared ~905nm
Total optical output power
LEDs less than 15mW
Sensor LED light emissions
Fall within Class 1 level, according to IEC 60825
-
1:2001
SpO
2
Measuring range
0
-
100%
Resolution
1%
Accuracy
At 70
-
100%, ±2%
At 0
-
69%, unspecified
Data update period
<13 seconds
Pulse Rate
Measuring Range
25
-
250bpm
Resolution
1bpm
Accuracy
±1% or ±1bpm, whichever is greater
Data update period
<13 seconds
Power supply
2 x 1.5V AAA alkaline batteries
Type of electrical protection
Class II
IP protection level
IPX1
Regulations applicable
Electrical Safety IEC 60601
-
1
Electromagnetic Compatibility IEC 60601
-
1
-
2
Electromagnetic Emissions
RF emissions CISPR 11
Group 1, uses RF energy only for its internal function. RF emis-
sions very low.
RF emissions CISPR 11
Class B, suitable for all establishments, including domestic use.
Environmental Conditions
Storage and Transport conditions
Temperature: Min
-
20°C, Max +55°C
Atmospheric Pressure: 500hPa
-
1060hPa
Humidity:
≤93%
RH
Operating conditions
Temperature: Min +5°C, Max +40°C
Atmospheric Pressure: 700hPa
-
1060hPa
Humidity:
≤85%
RH