4-22
4.2 Operation and features
SPM-32A ... SPM-36A
4.2.2
Inputs, input impedances (Z/
Ω
)
Note:
The instrument has various input impedance options;
the 150
Ω
input impedance can be replaced by
impedances of 120, 124, or 140
Ω
. The following
deals with the 150
Ω
version of the instrument, but
applies in principle to the other versions.
With the instrument you have the choice of 2 selectable
inputs:
•
Coaxial input:
Input impedances: 75
Ω
or
∞
(> 6.5 k
Ω
).
•
Balanced input
Input impedances: 75
Ω
, 150
Ω
, 600
Ω
or
∞
(>13 k
Ω)
.
Only one input should be selected at any one time
Selecting the input and the input impedance
The input and the input impedance are selected together:
»
WARNING
+30 dBm
«
The input level (a.c+d.c)
must not exceed
1W
or
+30 dBm
Max. d.c. input voltage=
60 V
from Z
out
≥
600
Ω
2
1
Z/
∞
BAL/UNBAL
3
75
600
150
150