Operating Instructions DMMS_DIGISYS
61
Set up measuring section / Calibration manual DMMS DIGISYS
1st measuring:
mean value: 475digits
laboratory sample: 5.0%
15. As soon as further moisture values become available, repeat the
"Take Sample" procedure.
2nd measuring: mean value: 528 digits
laboratory sample:
7,3%
16. Switch to "Curves" register card, carry out „...Add New Material“
and designate a material name.
17. Select "Type, 2..10 points".
18. Type in the values 475 and 5.0 into the fields Digits / M% carry
out "...Add Point".
19. Type in the values 528 and 7.3 into the fields „Digits“ and „M%“
carry out "...Add Point".
You can add from 2 to 10 points to one curve.
20. Select the old points and carry out "...Delete Point" respectively.
The new curve will also be displayed in the graphic.
21. If everything is OK, carry out "... Take Over Curve".
22. Confirm the message with "Yes".
23. Store new curve.
24. Confirm the measuring results by way of laboratory samples!