Operating Instructions DMMS_DIGISYS
59
Set up measuring section / Calibration manual DMMS DIGISYS
9.
Set up measuring section / Calibration manual DMMS DIGISYS
9.1
Measuring section
Sand is transported via a belt feeder from silo to the production.
The height of the material on belt feeder is constant. The on-line moisture measuring will be realised with
ACO moisture measuring sensor and slide. The moisture measuring sensor is integrated into the slide. They
are on top on the sand and both slides over/through the sand.
The moisture measuring sensor has the bus address “1”.
The 1
st
analogue output should be 4…20mA for a measuring range from 0…20% moisture.
9.2
Commissioning
1. At first you have to plug the delivered WAGO Service Cable (750-
920) or WAGO USB Service Cable (750-923/000-001) between
evaluation unit DIGISYS and PC/laptop for getting a connection.
2. Start the DIGISYS program and click on "Connect" in the icon
bar.
3. If the module is recognised, click on „YES“