7.8.
How To
–
Replace the Sample Port
–
Dust Filter
Remove Sample Port
–
Protection Cap
1.
Remove the protection cap by pulling it
sideward.
Replace Sample Port
–
Dust Filter
1.
Unscrew the dust filter counter-clockwise.
2.
Replace the dust filter by a new one.
3.
Replace the protection cap.
Calibrate the device after installation of a new dust filter.