Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 15 of 60
Calibrate the device after installation of a new filter cartridge.
4.4.
Sample Port
–
Installation
How To
–
Install the Sample Port
–
Dust Filter
1.
Remove the sample port
–
protection cap by pulling sideward.
2.
Check that the sample port
–
dust filter is assembled and free of impurities.
3.
Replace sample port
–
protection cap with gas intake hole downward.
Calibrate the device after installation of a new dust filter.
CAUTION
Do not use the device without filter cartridge, dust filter and protection cap. Impurities
within ambient air can get into gas path and destroy the sensor.