1. Introduction 

 

 

 

 

 

 

 

 

 

 

- 2

 

 

 

1-1 Introduction 

 

 

 

Model ACE-1000 is a VW data recorder, designed to measure a VW sensor, which 
generates a frequency with designed as a 4.8VDC low current charging circuit. 
 
The ACE-1000 can store up to 4,000 measurement values for up to 64 data sets, and this 
data can be transmitted to the computer via RS-232 communication, so it is useful in a large 
field and the software has data sorting function It is possible to sort by sensor ID, date, and 
time. 
 
The ACE-1000 adopts 256 × 128 graphic large LCD to increase the visibility, it displays the 
measurement related information at the same time and it is easy to use, and the backlight 
function is installed so that it can be easily measured at night on the spot. On the large LCD, 
the basic unit of the vibration sensor and the selected sweep band are displayed by default, 
and the selected units (10

3

Hz

2

, 4 types of strain, μsec) under Hz unit with displaying the 

thermistor RTD temperature sensor selected from thermistor also the power, current time 
are displayed simultaneously. 
 
Among the six keys, the display unit and frequency band can be selected by the menu key. 
If you select the frequency band, you can increase the resolution of the output device and 
perform precise measurement. 
 
Function to automatically shut off the power by user selection up to 60 minutes in 10 
minutes to efficiently extend the battery life and the usage period, and it can measure 30 
hours continuously with the Ni-MH rechargeable battery. You can use 10 years without 
replacement. Another key feature is the ability to adjust the LCD resolution so that it can be 
accurately measured in cold or hot areas. 
 
Impact resistant ABS Injection molding case is used for waterproof and damp proofing, and 
it can be used in any bad condition. It is a high reliability product with advanced electronic 
technology. It has better performance and function than any existing product.

 

 

1-2 Feature & Advantages   

 

 

 

Model ACE-1000 has features as following: 
 

 Waterproof, damp-proof, portable structure 

 High accuracy and resolution 

 Adoption of Ni-MH rechargeable battery 

 Usable in low temperature 

 With Back Light function, possible to measure at night 

 LCD brightness function 

 Auto power off function 

 Display battery lamp (Charge mark: LED)

 

 

1-3 Main function 

 

[ON]    Turn on the power. 

     

[OFF]    Turn off the power. 

     

[MENU]    Choose the level of Sweep or Unit, Contrast and temperature sensor. 

     

[ENTER]    It is for changing the sweep or unit, temperature sensor and contrast. 

     

[▲/LIGHT] 

 

It is for changing the sweep or unit, temperature sensor and contrast, and for turning on and 
off the backlight. 

     

[▼/SAVE]    It is for changing the sweep or unit, temperature sensor and contrast, and for saving data. 

Summary of Contents for ACE-1000

Page 1: ...Warranty VW Data Recorder Model ACE 1000...

Page 2: ...akdown owing to carelessness has happened and other operations such as dealing with zero point are needed services would be provided at a minimum cost 3 But in the case the purpose of use is changed i...

Page 3: ...g 5 3 1 Detail description for the function 5 3 2 Description of MENU 6 3 3 How to use ACE 1000 7 4 Program usage 8 4 1 Connect ACE 1000 to computer 9 4 2 A1000Pro Connect 9 4 3 A1000Pro Edit 10 4 4 A...

Page 4: ...measurement Function to automatically shut off the power by user selection up to 60 minutes in 10 minutes to efficiently extend the battery life and the usage period and it can measure 30 hours conti...

Page 5: ...Setting mode of MENU Button Transducer Connector Connect to a jumper cable to measure the frequency RS 232 Connector It is useful to download or delete the saved data from the computer or to set the...

Page 6: ...Communication port RS 232 Port Temperature sensor Thermistor 3k RTD 2k Temperature resolution 0 1 Display 256 128 Graphic LCD Keyboard Membrane key 6 Function keys pad Data capacity 4 000 Data points...

Page 7: ...isplay Mode 2 103 Hz2 2 Click Display Mode 3 0 391 3 Click Display Mode 4 0 7756 4 Click Display Mode 5 3 304 5 Click Display Mode 6 4 062 1 Click LCD Contrast 10 LCD Brightness adjustment 2 Click POW...

Page 8: ...ease Click ID change decrease 1 Click Saving measured data 1 Click Return to main display Reference Used while the measured value is on the screen Transducer Connector 10pin Jumper Cable Connect 10Pin...

Page 9: ...Applying sensors Mode 1 Period sec s All VW Sensors Mode 2 103 Hz2 103 Hz2 VW Load Cells Mode 3 Strain 0 391 0 391 VW Spot weldable Strain Gage Mode 4 Strain 0 7756 0 7756 VW Shotcrete Strain Gage Mod...

Page 10: ...ike MS Excel Only Ace Instrument offers A1000Pro Application description The Application program belonging with our VW data recorder ACE 1000 named A1000pro exe can be also downloaded from our website...

Page 11: ...he shape of follow ACE 1000 Connect If operate the program display main screen as follows Icon of Edit and Down is not activated because this is not connect For connect of VW data recorder turn on the...

Page 12: ...ow that able to check for battery of data recorder and set an option of Clock setup Reading setup Save option Auto power break Clock setup The first you must set time of computer and VW data recorder...

Page 13: ...4 Program usage 5 11 Reading setup Reading setup menu can be remote control the Sweep mode and Temperature sensor type ACE 1000 Sweep mode change ACE 1000 Temp Sensor Type change...

Page 14: ...rding mode the oldest data is deleted and the latest data is saved when the save capacity is full If you choice Stop recording mode subsequent data is not save when the save capacity is full Auto powe...

Page 15: ...00Pro Down If click the Down icon in the connected screen display the screen that arranged all saved data in VW data recorder as follows When a screen listing all saved data appears click ID option it...

Page 16: ...ion of csv and dat for using in operation program like MS Excel And this make use of extension of txt for checking in text file program like WordPad After choose the txt file save it Then WordPad prog...

Page 17: ...ck the storage space and delete the data If you click the Erase button on the Down screen a question about full deletion will be displayed After confirmation all data will be deleted 4 5 A1000Pro Exit...

Page 18: ...with great force When not using the jumper cable put it in a bag or case and store it in a place where it will not be damaged Check battery voltage Confirm the voltage of the built in battery before o...

Page 19: ...ce from the measuring machine an amplifier should be installed in proportion of measurement distance to prevent drop of voltage for sensor transmission 8 Atmospheric pressure adjustment When manometer...

Page 20: ...5 17...