ACE INSTRUMENTS ACE-1000 Instruction Manual Download Page 19

 

6. Geotechnical Instruments Installation and Operation Tips 

 

 

 

 

 

 

 

 

 

- 17 

 

 

 
1. Handling caution 

Engineering measuring instrument is precisely 

manufactured and sensitive. Don

’t drop it and don’t allow it 

to be exposed to external shocks. In particular, the VW 
measuring instrument is likely to see its zero point changed 
due to shocks. 

 

2. Caution for storing calibration report 

Calibration report is provided with each sensor. 

Information at the moment of calibration includes 
barometric pressure, calibration temperature, and 
temperature sensor in use, calibration data, conversion 
ratio and classification of signal cable by color. Therefore, 
you should be careful with storing the information until all 
work is done. In case the report is lost, tracking data and 
product might be impossible. 

 

3. Operation by an engineer 

All the process should be operated by a skilled engineer to 

prevent inappropriate choices such as errors of instrument 
choice, installation and operation which make impossible 
near-permanent calculation. 

 

4. The need for shielding 

It is common that sensor sends a weak electronic signal. 

And sensor is under the influence of electromagnetic 
induction, electrostatic induction static electricity 
electrification generated when other control machine is on. 
In particular, given that environmental condition of 
engineering work involving welding, generator, motor, 
antenna, and watery surface is poor, shielding and ground 
wire should be connected in the case of output device 
connection. 

 

5. Caution for signal cable extension 

There would be no problem that VW sensor outputting 

frequency signal is connected for extension .in the case 
that a sensor generating Voltage is connected for 
extension, electronic sensors are vulnerable to changes in 
resistance value.    Therefore, the length of the cable plays 
a big influence. So, you should adjust resistance change 
value with adjustment value toward a sectional length of 
signal cable by manufacturers. Also, extension seam part 
should be finished up with Splice Kit (Epoxy). 

 

6. The use of protection tube for signal cable 

In the case of laying under the ground, dams, vulnerable 

foundation, concrete structures can have a big 
displacement, a source of signal cable disconnection. In 
the case of ground construction, given the construction 
condition, heavy equipment operation and frequent moving 
construction can be a source of disconnection or sensor 
damage. Therefore, caution should be taken for the 
protection of sensors and signal cables. And minimize the 
dangerous factors by using new construction pipe for 
drainage system. 

 

7. Installation of Amplifier with sensors generating mV 

Engineering  measuring  machine  generating  electronic 
signal has a 200-300 m transmission distance. In the case 
Junction Box and Terminal Box have a long distance from 
the measuring machine, an amplifier should be installed in 
proportion  of  measurement  distance  to  prevent  drop  of 
voltage for sensor transmission.

 

 

 
8. Atmospheric pressure adjustment 

When manometer is chosen as sensor, it is very sensitive 

to change in atmospheric pressure. Therefore, for precise 
measurement, places such as dams, valleys, the sea and 
reclaimed lands with high fluctuation of atmospheric 
pressure should adjust head height according to the 
difference of the pressure by using mercury barometer. 

 

9. Lighting rod construction 

In the case of large engineering work near water such as 

dams, lakes, the sea and large fields, valleys, lighting rods 
connecting each sensor individually should be constructed 
to protect sensors. Within the influence of over voltage, 
sensors are likely to function poorly. 

 

10. Caution for filter use in pressure sensors 

Pressure sensors such as negative pore water pressure 

have attached filters. Upon installation, make sure to let air 
out of the inside of the filter. Given that densities of air 
contraction and water are different, remained air might 
cause an error to measured value. Also, at places where 
unsaturated soil or negative pore water pressure are 
expected, it is recommended to use 1

μ

m high-density 

ceramic filter. 

 

11. Caution for bearing plate in use for load cell 

For maintaining consistent measured value and high 

precision, the use of bearing plate is important, during 
installation of hard cell on Earth Anchor, heat processed 
steel materials should be manufactured with the enough 
thickness to endure unloading load and with the density of 
within 0.5 degree in top-down parallel lines. High quality 
products should be used for cone for spilt, cone-plate and 
mutual taper processed density. 

 

12. Sensor temperature adjustment 

An element of VW sensors uses wire rod (used for piano 

string) so that coefficient of linear expansion metallic 
materials hold can cause errors of margin and sensors of 
electricity and electron hold an adjustment factor. For 
precise measurement, adjust the difference of temperature 
with a mercury thermometer. 

 

13. Auto Data Acquisition System requires UPS 

In South Korea, relatively temporary blackout (0.5 

seconds) is frequent. So overloading owing to simultaneous 
uses of equipment and accumulated power cables on sites 
are likely to cause ordinary and temporary blackouts, 
leading to computer Down and errors of built-in software.   
Therefore, When Auto Data Acquisition System is operated; 
UPS(Uninterruptable Power Supply) must be used to 
minimize dangerous factors. 

 

14. Caution for the choice of VW output unit 

Manufactures of VW sensors and output units usually cover 
600~3,200Hz  (360~10,240  10

3

Hz

2

 

or  1666

312 

μsec)  for 

possible  measurement  range.  And  they  design  sensors 
taking into consideration the purpose of sensors, durability, 
and  precision.  Therefore,  when  Mode  for  unit  choice  of 
output  unit  within  this  range  is  changed,  you  must  select 
sophisticated output unit-making measurement possible

 

Summary of Contents for ACE-1000

Page 1: ...Warranty VW Data Recorder Model ACE 1000...

Page 2: ...akdown owing to carelessness has happened and other operations such as dealing with zero point are needed services would be provided at a minimum cost 3 But in the case the purpose of use is changed i...

Page 3: ...g 5 3 1 Detail description for the function 5 3 2 Description of MENU 6 3 3 How to use ACE 1000 7 4 Program usage 8 4 1 Connect ACE 1000 to computer 9 4 2 A1000Pro Connect 9 4 3 A1000Pro Edit 10 4 4 A...

Page 4: ...measurement Function to automatically shut off the power by user selection up to 60 minutes in 10 minutes to efficiently extend the battery life and the usage period and it can measure 30 hours conti...

Page 5: ...Setting mode of MENU Button Transducer Connector Connect to a jumper cable to measure the frequency RS 232 Connector It is useful to download or delete the saved data from the computer or to set the...

Page 6: ...Communication port RS 232 Port Temperature sensor Thermistor 3k RTD 2k Temperature resolution 0 1 Display 256 128 Graphic LCD Keyboard Membrane key 6 Function keys pad Data capacity 4 000 Data points...

Page 7: ...isplay Mode 2 103 Hz2 2 Click Display Mode 3 0 391 3 Click Display Mode 4 0 7756 4 Click Display Mode 5 3 304 5 Click Display Mode 6 4 062 1 Click LCD Contrast 10 LCD Brightness adjustment 2 Click POW...

Page 8: ...ease Click ID change decrease 1 Click Saving measured data 1 Click Return to main display Reference Used while the measured value is on the screen Transducer Connector 10pin Jumper Cable Connect 10Pin...

Page 9: ...Applying sensors Mode 1 Period sec s All VW Sensors Mode 2 103 Hz2 103 Hz2 VW Load Cells Mode 3 Strain 0 391 0 391 VW Spot weldable Strain Gage Mode 4 Strain 0 7756 0 7756 VW Shotcrete Strain Gage Mod...

Page 10: ...ike MS Excel Only Ace Instrument offers A1000Pro Application description The Application program belonging with our VW data recorder ACE 1000 named A1000pro exe can be also downloaded from our website...

Page 11: ...he shape of follow ACE 1000 Connect If operate the program display main screen as follows Icon of Edit and Down is not activated because this is not connect For connect of VW data recorder turn on the...

Page 12: ...ow that able to check for battery of data recorder and set an option of Clock setup Reading setup Save option Auto power break Clock setup The first you must set time of computer and VW data recorder...

Page 13: ...4 Program usage 5 11 Reading setup Reading setup menu can be remote control the Sweep mode and Temperature sensor type ACE 1000 Sweep mode change ACE 1000 Temp Sensor Type change...

Page 14: ...rding mode the oldest data is deleted and the latest data is saved when the save capacity is full If you choice Stop recording mode subsequent data is not save when the save capacity is full Auto powe...

Page 15: ...00Pro Down If click the Down icon in the connected screen display the screen that arranged all saved data in VW data recorder as follows When a screen listing all saved data appears click ID option it...

Page 16: ...ion of csv and dat for using in operation program like MS Excel And this make use of extension of txt for checking in text file program like WordPad After choose the txt file save it Then WordPad prog...

Page 17: ...ck the storage space and delete the data If you click the Erase button on the Down screen a question about full deletion will be displayed After confirmation all data will be deleted 4 5 A1000Pro Exit...

Page 18: ...with great force When not using the jumper cable put it in a bag or case and store it in a place where it will not be damaged Check battery voltage Confirm the voltage of the built in battery before o...

Page 19: ...ce from the measuring machine an amplifier should be installed in proportion of measurement distance to prevent drop of voltage for sensor transmission 8 Atmospheric pressure adjustment When manometer...

Page 20: ...5 17...

Reviews: