Patient Image Analysis
Image Enhancement Gammas
Linear
The Linear gamma represents a "standard" processing
algorithm. Linear describes a greyscale representation
of a B-Scan Plus echogram in which the brightness
represented on the screen is directly proportional to the
strength of the ultrasonic echo. Therefore an ophthalmic
structure which reflects the maximum echo would be
represented as pure white, while no echo would be
represented as pure black; and echoes in between
would be represented proportionally as shades of gray,
directly proportional to the strength of the ultrasonic
echo. Figure 119 below displays an example of a B-
Scan Plus image with a Linear Full Range gamma
setting.
Figure 119
Linear Full Range Gamma Setting
Exam Info
Current View
Probe Controls
Scan Controls
All Views
TGC Adjustments
Gamma Adjustments
f
Zoom Controls
Summary of Contents for B-Scan Plus
Page 1: ...Accutome B Scan Plus User s Guide 24 6110 Rev D...
Page 20: ...xviii Accutome B Scan Plus User Guide...
Page 26: ...6 Accutome B Scan Plus User Guide...
Page 54: ...34 Accutome B Scan Plus User Guide Figure 16 TGC Tab Active Figure 17 Gamma Tab Active...
Page 108: ...88 Accutome B Scan Plus User Guide...
Page 114: ...94 Accutome B Scan Plus User Guide Figure 74 Initial Exam Screen with Patient Name...
Page 173: ...153 Patient Image Analysis TGC Adjustment Figure 118 TGC Adjustments Displayed...
Page 192: ...172 Accutome B Scan Plus User Guide...
Page 196: ...176 Accutome B Scan Plus User Guide Figure 136 Printing a Frame PRINT IMAGE...
Page 209: ...189 Printing Images and Reports Creating Reports...
Page 210: ...190 Accutome B Scan Plus User Guide...
Page 254: ...234 Accutome B Scan Plus User Guide...