4.5.2.5
The
MENU
,
,
,
and
keys can be used
together to adjust these parameters. The parameters can
also be saved into the instrument. The following is the
procedure of adjusting and saving the parameters of Rtp
and a4 for an SPRT type probe. The other parameter
adjustments follow the same process.
4.5.2.6
To adjust Rtp (continuing from step 4.5.2.4) press
or
key to move the cursor to “>SET Rtp”, and then
press
key to go into the Rtp adjustment program.
[SENSOR 1 SPRT]
MENU=Exit <^v> = Adj
Rtp = OHM
25.5260
The instrument will display the current Rtp value. The digit
underlined with the blinking bar indicates the adjustable
position. Users can use
and
keys to move this
position and use
and
keys to adjust the value.
4.5.2.7
After the Rtp is adjusted, press
MENU
key, the display will
remind user to save the adjusted value by flashing the
adjusted Rtp on screen. Press
key to save the value or
press
MENU
key to exit the program without saving and go
back to the parameter adjustment menu.
4.5.2.8
The following is a4 adjustment procedure (Continuing
from step 4.5.2.4). Pressing
and
keys to move the