![Acculab ALC series Operating Instructions Manual Download Page 9](http://html1.mh-extra.com/html/acculab/alc-series/alc-series_operating-instructions-manual_3767675009.webp)
Calibration/Adjustment*
Always calibrate/adjust the balance/scale after setting it up in a new location.
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance/scale,
– the balance/scale is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed.
The weight required for calibration/adjustment is displayed (see “Accessories” for calibration weights).
On ALC balances, calibration/adjustment can be performed with any of three weight units: g, kg**, or lb (menu code
1. 4.
x)
Factory setting: Grams (menu code
1. 4. 1
)
Example
Calibrate the balance/scale
Menu code settings: Factory-set codes
Step
Key (or instruction)
Display/Printout
1. Turn on the balance/scale
[ON/OFF]
2. Tare the balance/scale
[ZERO]
0.0
g
3. Begin calibration
[CAL]
+
1000.0
g
Calibration weight is displayed
without weight unit (here: 1000 g)
4. Place the indicated calibration
1000.0
g
weight on the balance/scale
After calibration, the calibration
+
1000.0
g
weight is displayed with wt. unit
5. Remove the calibration weight
0.0
g
* = “Calibration” technically means to determine the difference between the balance/scale readout and the actual weight
on the pan to determine the accuracy. Adjustment means to bring a balance/scale into the state of accuracy required for its use.
Therefore, “calibration,” as used in this manual, actually means “adjustment.”
**= not on models with a readability of 0.1 mg or 1 mg
9
www.balances.com is your Acculab deep discount distributor