![Acculab ALC series Operating Instructions Manual Download Page 22](http://html1.mh-extra.com/html/acculab/alc-series/alc-series_operating-instructions-manual_3767675022.webp)
Error Codes
Error codes are shown on the main display for approx. 2 seconds, after which the program automatically returns to the weighing mode.
Display
Cause
Solution
No segments appear
No AC power is available
Check the AC power supply
on the display
The AC adapter is not plugged in
Plug in the AC adapter
Battery is dead
Replace the battery
Recharge the battery using an external charger
h
The load exceeds the balance/scale capacity
Unload the balance/scale
l
The weighing pan is not in place
Place the weighing pan on the balance/scale
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
e 01
Display capacity exceeded:
Decrease the weight on the balance/scale
Value to be output cannot
be shown on the display
e 02
Calibration parameter
Calibrate only when zero
requirement not met; e.g.:
is displayed
– balance/scale not zeroed
Press [ZERO] to zero the balance/scale
– balance/scale is loaded
Unload the balance/scale
e 09
When gross value
<
zero, no tare
Tare the balance/scale
e 10
The [ZERO] key is blocked when
Press [CAL] to clear the tare memory and release
there is data in the second tare
the tare key
memory (net-total) – only 1 tare
function can be used at a time
e 11
Value input is not allowed for
Press [ZERO]
second tare memory
e 22
Weight is too light or there is
Increase the reference quantity
no sample on the balance/scale
or sample amount
e 30
Interface port for printer output is blocked
Contact your local Acculab Service Center
Max. weighing capacity
The balance/scale was
Place the weighing pan on the
is less than indicated
turned on without the
balance/scale and press [ON/OFF] to
under “Specifications”
weighing pan in place
turn the balance/scale back on
The weight readout
The balance/scale has not
Calibrate/adjust the balance/scale
is obviously wrong
been calibrated/adjusted
The balance/scale was not
Tare before weighing
tared before weighing
If any other errors occur, please contact your local Acculab Service Center.
22