![Acculab ALC series Operating Instructions Manual Download Page 10](http://html1.mh-extra.com/html/acculab/alc-series/alc-series_operating-instructions-manual_3767675010.webp)
Application Programs
Net-Total Formulation/Second Tare Memory
With this application program you can weigh in components for formulation of a mixture.
Preparation
Configure the Net-Total Formulation/Second Tare Memory application in the operating menu: See “Configuring the Balance/Scale”
Menu code
2.1.3
Example
Step
Key (or instruction)
Display/Printout
1. Turn on the balance/scale
[ON/OFF]
2. Place an empty container on
+
65.0
g
the balance/scale
3. Tare the balance/scale
[ZERO]
4. Add the first component
+
120,5
g
5. Store the first component weight
[F]
0.0
g
NET
If the print format is set
to include data ID codes,
the following is printed:
N1 + 120.5 g
6. Add the next component
+
70,5
g
7. Store the 2nd component weight
[F]
0.0
g
NET
8. Add further components,
As described
if desired
for steps 5 and 6
9. Display total weight
[CAL] <2 sec
+
191,0
g
and fill to desired final weight
10