113
Acuvim-L
Response
The normal response to the command request is to retransmit the message as
received after the Relay status has been altered.
Table 4-23 Control Relay Response Message
Transaction
identifier hi
Transaction
identifier lo
Protocol
identifier hi
Protocol
identifier lo
Length hi Length lo
Unit
identifier
00H
00H
00H
00H
00H
06H
01H
Fun
Data start reg hi
Data start reg lo
Value hi
Value lo
05H
00H
00H
FFH
00H
e. Preset/Reset Multi-Register (Function Code 16)
Query
Function code 16 allows the user to modify the contents of a Multi-Register. The
example below is a request to an Acuvim-L series meter address 1 to Preset CT1
(500) and CT2 (5). CT1 data address is 108H, and CT2 data address is 109H.
Table 4-24 Preset CT Value Query Message
Transaction
identifier hi
Transaction
identifier lo
Protocol
identifier hi
Protocol
identifier lo Length hi Length lo
Unit
identifier
00H
00H
00H
00H
00H
0BH
01H
Fun Data start
reg hi
Data start
reg lo
Data #of
regs hi
Data #of
regs lo
Byte
count
Value1
hi
Value1
lo
Value2
hi
Value2
lo
10H
01H
08H
00H
02H
04H
01H
F4H
00H
05H
Response
The normal response to a preset Multi-Register request includes MBAP Header,
function code, data start register and the number of registers.