IMAGE PROBLEMS 

Problem 

Cause 

Corrective Measures 

Image moves while focusing 

Specimen rises from stage 
surface 

Revolving nosepiece is not in 
the click-stop position 

Secure the specimen in the 
slide holder 

Revolve the nosepiece to the 
click-stop position 

Image tinged yellow 

Lamp intensity is too low  

 

 

 

Blue filter not used 

Adjust the light intensity by 
rotating the intensity control 
dial and/or iris diaphragm 

 

Use daylight blue filter 

Image is too bright 

Lamp intensity is too high 

Adjust the light intensity by 
rotating the intensity control 
dial and/or iris diaphragm 

Insufficient brightness 

Lamp intensity is too low 

 

Aperture diaphragm closed too 
far 

Condenser position too low 

Adjust the light intensity by 
rotating the intensity control 
dial and/or iris diaphragm 

Open to the proper setting 

 

Position the condenser slightly 
lower than the upper limit 

  

  

 

  

  

  

MECHANICAL PROBLEMS 

Image will not focus with high 
power objectives 

Slide upside down 
 

 

Cover glass is to thick 

Turn the slide over so the 
cover glass faces up 

 

Use a 0.17mm cover glass 

High power objective contacts 
slide when changed from low 
power objective 

Slide upside down  
 

 

Cover glass is to thick 

Diopter adjustment is not set 
properly 

Turn the slide over so the 
cover glass faces up 

 

Use a 0.17mm cover glass 

 

Readjust the diopter settings 
as outlined in section 4.3 

Summary of Contents for EXM-150 SERIES

Page 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com EXM 150 MICROSCOPE SERIES...

Page 2: ...CE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS DIAGRAMS 4 5 ADJUSMENT OPERATION CONNECTING THE POWER CORD 6 ILLUMINATION 6 SPECIMEN SLIDE 7 FOCUSING 7 DISC OR IRIS DIAPHRAGM 8 OIL OBJECTIVE 8 RECHARGING...

Page 3: ...ponent including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More pers...

Page 4: ...ighest quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived pre assembled and packed in a molded container Do not discard the container the molded container should be retained fo...

Page 5: ...25 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Other Models Not Shown EXM 150 MS EXM 151 MST EXM 150 T EXM 151 MS Y Axis Adjustment Top Knob X Axis Adjustment Bottom Knob Condenser Models Fea...

Page 6: ...pe uses a universal power supply so it may be used in any voltage range between 90 240v when used with the appropriate line cord IMPORTANT Use care when storing the power cord so that it does not bend...

Page 7: ...a specimen slide into the slide holder with the cover glass facing up Release the slide finger so it closes and secures the slide in place Rotate the X and Y axis knobs to move the slide into positio...

Page 8: ...cimen slide Rotate the nosepiece so the 100xR oil immersion objective is in the light path Dust or air bubbles in the oil can destroy the definition of the image If the bubbles are trapped between the...

Page 9: ...000A line cord charger and requires approximately 8 hours for a full charge The life expectancy of the rechargeable batteries is approximately 20 000 hours or equivalent to 400 recharges before needin...

Page 10: ...ies 6 Slide the cover back on and replace the screw 7 Close the trap door and tighten the lock screw 8 Carefully place the microscope in an upright position 9 Plug the charger to the power receptacle...

Page 11: ...ck stop position by swinging the objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the len...

Page 12: ...rotating the intensity control dial and or iris diaphragm Insufficient brightness Lamp intensity is too low Aperture diaphragm closed too far Condenser position too low Adjust the light intensity by...

Page 13: ...oes not light when switched on No electrical power Lamp bulb burnt out Fuse blown out Check power cord connection Replace bulb Replace fuse Slippage of focus when using the coarse focusing knob Tensio...

Page 14: ...orkmanship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of two years from the date of invoice to the original end user...

Reviews: