TROUBLESHOOTING 

Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a 
problem occurs, please review the following list and take remedial action as needed. If you cannot solve the 
problem after checking the entire list, please contact your local dealer for assistance. 

 

OPTICAL 

Problem 

Cause 

Corrective Measure 

Darkness at the periphery or 
uneven brightness of view field 

Revolving nosepiece not in 
click stop position 

Rotate the nosepiece to click 
stop position by swinging the 
objective correctly into the 
optical path 

Dirt or dust on the view field 

Dirt or dust on the lens - 
eyepiece, condenser, objective, 
collector lens or specimen 

Clean the lens 

Poor image quality 

No cover glass attached to the 
slide  
 
Cover glass is too thick or thin 
 
 
 
Slide maybe upside down 
 
 
Immersion oil is on a dry 
objective (especially the 40xR) 
 
No immersion oil used 
with100xR objective 
 
Air bubbles in immersion oil 
 
Condenser aperture is closed 
or open too much 
 
Condenser is positioned too 
low 

Attach a 0.17mm cover glass 
 
 
Use a cover glass of the 
appropriate thickness 
(0.17mm)  
 
Turn slide over so the cover 
glass faces up 
 
Check the objectives, clean if 
necessary 
 
Use immersion oil 
 
 
Remove bubbles 
 
Open or close properly 
 
 
Position the condenser slightly 
lower than the upper limit 

  

 

 

Summary of Contents for EXM-150 SERIES

Page 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com EXM 150 MICROSCOPE SERIES...

Page 2: ...CE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS DIAGRAMS 4 5 ADJUSMENT OPERATION CONNECTING THE POWER CORD 6 ILLUMINATION 6 SPECIMEN SLIDE 7 FOCUSING 7 DISC OR IRIS DIAPHRAGM 8 OIL OBJECTIVE 8 RECHARGING...

Page 3: ...ponent including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More pers...

Page 4: ...ighest quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived pre assembled and packed in a molded container Do not discard the container the molded container should be retained fo...

Page 5: ...25 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Other Models Not Shown EXM 150 MS EXM 151 MST EXM 150 T EXM 151 MS Y Axis Adjustment Top Knob X Axis Adjustment Bottom Knob Condenser Models Fea...

Page 6: ...pe uses a universal power supply so it may be used in any voltage range between 90 240v when used with the appropriate line cord IMPORTANT Use care when storing the power cord so that it does not bend...

Page 7: ...a specimen slide into the slide holder with the cover glass facing up Release the slide finger so it closes and secures the slide in place Rotate the X and Y axis knobs to move the slide into positio...

Page 8: ...cimen slide Rotate the nosepiece so the 100xR oil immersion objective is in the light path Dust or air bubbles in the oil can destroy the definition of the image If the bubbles are trapped between the...

Page 9: ...000A line cord charger and requires approximately 8 hours for a full charge The life expectancy of the rechargeable batteries is approximately 20 000 hours or equivalent to 400 recharges before needin...

Page 10: ...ies 6 Slide the cover back on and replace the screw 7 Close the trap door and tighten the lock screw 8 Carefully place the microscope in an upright position 9 Plug the charger to the power receptacle...

Page 11: ...ck stop position by swinging the objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the len...

Page 12: ...rotating the intensity control dial and or iris diaphragm Insufficient brightness Lamp intensity is too low Aperture diaphragm closed too far Condenser position too low Adjust the light intensity by...

Page 13: ...oes not light when switched on No electrical power Lamp bulb burnt out Fuse blown out Check power cord connection Replace bulb Replace fuse Slippage of focus when using the coarse focusing knob Tensio...

Page 14: ...orkmanship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of two years from the date of invoice to the original end user...

Reviews: