3075-GS ZOOM STEREO MICROSCOPE ON GEM STAND
ACCU-SCOPE
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
10
REPLACING THE LAMPS
Replacing the Halogen Lamp in the
Bottom Illuminator (Fig. 11)
The darkfield illuminator (bottom light) uses
one 6 volt, 20 watt tungsten halogen lamp –
CAT #6047C
.
To replace the lamp, tilt the Microscope all the
way back (see page 8).
On the bottom of the darkfield housing is a trap
door. Open the trap door by loosening the
screw (see Fig. 11). The screw can be loosened
by hand or by using the small flathead
screwdriver included in the original shipment of
the Gem stand. The door will open and the
lamp can be removed.
IMPORTANT:
When replacing the lamp, be
careful not to touch the new lamp with your
fingers as the oils and dirt on your fingers can
affect the lamp’s performance. To insert the
new lamp wrap the plastic bag supplied with the
lamp or a piece of lens tissue around the lamp
before handling/inserting it.
Fig. 11
Fig. 11