background image

Page 5 of 7   200.02 Point of Care ACCU-CHEK Inform ll Blood Glucose Monitoring System 

 

a.  The  date,  time,  name  of  person  contacted  must  be  documented  in 

the  computer  in  the  significant  events  form  including

 

documentation 

that read back verify (RBV) was done. 

ii. 

In  addition,  the  operator  should  send  a  specimen  to  the  laboratory  for  further 
quantification. 

iii. 

The  operator  should  always  follow  the  protocol  for  hypoglycemia  or 
hyperglycemia that is in place for the patient type. The operator should not wait 
for the lab confirmation before starting treatment. 
 

12.  Documentation 

a.  A licensed medical provider must order all patient testing. 
b.  Glucose results are documented in Powerchart 

i.  Labor and postpartum patients: ACCU-CHEK INFORM LL Inform II results must be 

charted under "Maternal Vital Signs” in QS. 

ii.  Well newborns in QS: ACCU-CHEK INFORM LL Inform II results must be charted on 

the newborn flow sheet under “glucose.”  

c.  If  patients  choose  to  recheck  results  with  their  own  glucose  meter,  these  results  cannot  be 

used for diagnostic purposes or be documented in the patient’s medical record.  

d.  If  a  patient  is  on  an  insulin  pump,  it  may  be  used  by  the  patient  if  the  licensed  medical 

provider  writes an order for the pump to be used. In this in

stance the patient’s own glucose 

meter may be used for glucose testing and insulin bol

using via the patient’s insulin pump. 

e.  Once results are verified the test will automatically charge in computer. 
f. 

If the value is outside for the reportable range (<30 or >500), the value entered in computer 
will be <30 or >500 with a comment added to indicate the actions taken.   

g.  The  glucose  value  of  the  patient  will  be  stored  in  the  meter  and  may  be  retrieved  at a  later 

date. 

h.  Whenever the patient result obtained by the ACCU-CHEK Inform ll does not appear to fit the 

clinical picture repeat the QC and patient specimen. 

A. 

You must use the comment “Repeat will not post” on the repeat test. The patient will 
not be charged for two tests. Only the repeat test will chart. 

 

13.  Troubleshooting 

a.  Troubleshooting  and  technical  assistance  with  the  ACCU-CHEK  Inform  ll  is  handled  by  the 

Point  of  Care  Coordinator  (POCC),  the  Education  office,  the  CRC,  superuser,  or  the 
superuser on the unit that is responsible for point-of-care glucose testing. 

c. 

Document the problem with the red “out of service” tags attached to the meter and bring the 
meter to the Laboratory. 

d.  Loaner meters are available. 
E.  Help  with  corrective  action  is  available  from  the  laboratory,  in  the  ACCU-Check  Inform  ll 

manual, and within 

the meter using the “Comment Codes”. 

 

13.  Storage of Testing Materials 

a.  Test strips should be stored at room temperature from 36-86°F. 
b.  Unopened test strips are good until the expiration date printed on the label. 
c.  When you take a test strip out of the container, put the cap back on tightly immediately. 
D.  Control material is good 90 days after opened if kept at 39-86°F (4-30°C). 

 

14.  Supplies 

a.  Strips  and  auto-disabling  single  use  fingerstick  lancets  are  stocked  in  Supply  Chain 

Management and stored in the PYXIS.. 

b.  Do not use the test strips if there is loose desiccant in the test strip vial. 
c.  Controls will be kept in the Laboratory and replaced every three months by the POCC.  

 

15.  Training and Proficiency 

a.  New operators need to show proficiency initially, at six months and yearly thereafter. This will 

be incorporated initially at orientation, at six months with the assistance of superusers and at 
the annual nursing competency review.  

b.  Glucose  proficiency  testing,  a  CAP  requirement,  will  be  performed  according  to  the 

scheduled distribution. 

c.  The testing material, directions and data sheets will be provided by the laboratory. 

Summary of Contents for Inform II Competency

Page 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Page 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Page 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Page 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Page 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Page 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Page 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Page 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Page 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Page 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Page 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Page 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Page 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Page 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Page 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Page 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Page 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Page 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Page 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Page 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Reviews: