background image

HP 300-121 

Page 2 of 5 

 

Reviewed/Revised 06/26/13 

Cleaning and Disinfecting the ACCU-CHEK Inform II System Components 

 

Meter and  Base  Unit  Cleaning  and  Disinfecting

:    Cleaning  and  disinfecting  the  exterior  surface  of 

the  meter  with  either  Clorox  Bleach  Wipes  is,  at  minimum,  recommended  daily  for  dedicated  patient 
devices.  Meters used with multiple patients may require more frequent cleaning and disinfecting. 

 

NOTE

:  Follow all facility safety and infection control policies when handling, cleaning and disinfecting 

ACCU-CHEK Inform II meters. 

QUALITY CONTROL 

Quality  control  testing  is  performed  as  a  primary  means  of  ensuring  ongoing  proper  performance  of  the 
ACCU-CHEK  Inform  II  system.    Low  and  high  quality  control  testing  is  performed  on  the  following 
occasions: 

1.  Every 24 hour period of patient testing. 

2.  When a new test strip vial is opened and placed into service. 

3.  With unexpected patient results of any kind. 

4.  Acceptable quality control results are defined as results that are within

 

manufacturer’s

 

posted limits. 

5. 

Quality  control  testing  results  must  be  acceptable  prior  to  performing  patient  testing, 
according to the 

Interpretation of Results

 section of the procedure below. 

Procedure 

1.  Turn on the ACCU-CHEK Inform II meter. 

2.  Enter  the  operator  ID  by  means  of  scanning  the  barcode  on  the  identification  band.   

NOTE:

    If  the 

operator ID you enter is not accepted, attempt to re-enter it.  If it is still rejected, contact the laboratory.

  

DO NOT 

attempt to perform tes

ts under another operator’s ID. 

3.  From the 

Main Menu

, touch

 Control Test

4.  Select the level of control that you wish to test. 

5.  Confirm that the meter is coded (calibrated) to the same test strip code that is printed on the test strip 

vial by scanning the test strip vial’s barcode. Contact the laboratory if unable to confirm the correct test 
strip code.

 

6.  The meter will display a picture of a test strip with a downward flashing arrow on the meter indicating 

that  you  are  ready  to  insert  a  test  strip  into  the  meter.    Remove  a  test  strip  from  the  vial  and 
immediately recap the  vial.   Insert the test strip  into the meter in the direction of the arrows and  with 
the “ACCU-CHEK” lettering facing upward.  The meter will display a flashing drop above the test strip 
icon when the test strip is properly inserted indicating that you are ready to apply control solution. 

7.  Apply  control  solution  to  the  front  edge  of  the  test  strip.    The  solution  will  fill  the  yellow  sample 

chamber by capillary action.  Do not apply sample to the top of the test strip.  Once sufficient sample 
has been detected, the measurement begins.  An hourglass icon indicates that the measurement is in 
progress.   You  will  get an  error message if the sample is insufficient.   If this occurs,  you  will need to 
repeat the test. 

8.  The  measurement  is  complete  when  the  result  is  displayed  on  the  meter  screen.    See  the 

“Interpretation  of  Results”  and  “Troubleshooting”  sections  below  for  guidance  on  what  to  do  if  your 
result is “Fail” or shows an “out of range” message associated with the result. 

9.  Remove the test strip and dispose of it in plain trash.

 

10.  Touch the comment button ( 

) to enter an appropriate comment(s) as required in the “General 

Policies” section of this manual.

 

11.  Touch the 

  button to confirm the result and send the result from the meter wirelessly or place 

the meter in the base unit to send the result and record the result into the electronic data management 
system. The base unit also charges the meter. 

Interpretation of Quality Control 

Summary of Contents for Inform II Competency

Page 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Page 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Page 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Page 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Page 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Page 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Page 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Page 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Page 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Page 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Page 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Page 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Page 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Page 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Page 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Page 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Page 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Page 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Page 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Page 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Reviews: