background image

Page 3 of 7   200.02 Point of Care ACCU-CHEK Inform ll Blood Glucose Monitoring System 

 

h.  Interpretation of Quality Control Results 

a.  Results are displayed on the screen as a number. 

Any result that shows an “out of range” 

message or “Fail” is an indication that the system may not be performing correctly for 
patient testing.

 

b.  Patient testing may not be performed if quality control testing results are not within 

acceptable limits and the meter will not display the patient testing option if scheduled 
quality control results exceed acceptable limits. 

 

9.  Special considerations for Patient Testing 

a.  Hematocrit should be between 10

–65 %.

 

b.  Lipemic samples (triglycerides) in excess of 1800 mg/dL may produce elevated results.

 

c.  Blood concentrations of galactose >15 mg/dL will cause overestimation of blood glucose 

results.

 

d.  Intravenous administration of ascorbic acid which results in blood concentrations of 

ascorbic acid >3 mg/dL will cause overestimation of blood glucose results. 

e.    If peripheral circulation is impaired, collection of capillary blood from the approved 

sample sites is not advised as the results might not be a true reflection of the 
physiological blood glucose level. This may apply in the following circumstances: severe 
dehydration as a result of diabetic ketoacidosis or due to hyperglycemic hyperosmolar 
non-ketotic syndrome, hypotension, shock, decompensated heart failure NYHA Class IV, 
or peripheral arterial occlusive disease. 

f. 

Hematocrit

—the acceptable range for the ACCU-CHEK INFORM ll test strips is 20-65% 

for values below 200 mg/dL and 20-55% for values above 200 mg/dL. 

g.  Patients in Isolation 

1.  The meter can be placed in a plastic bag. 
2.  A small hole can be punctured in the bag. 
3.  The strip can be placed in the meter through the hole and the meter programmed as 

usual. 

4.  After  the  testing  is  done  the  plastic  bag  is  discarded  and  the  meter  will  be  wiped 

down. 

 

10.  Procedure for Patient Testing 

a.  Wash  hands  and  don  personal  protective  equipment  (gloves,  gowns,  etc.)  as  required  by 

infection control and isolation policies and procedures. 

b.  Assemble the finger-stick equipment, the meter, and the test strips at the patient bedside. 
c.  Identify the patient by armband and explain the procedure. 
d. 

Wash  and  dry  patient’s  hands

/

heel  with  soap  and  water  before  obtaining  a  blood  sample. 

Alcohol wipes are not recommended for use.  

e.  Turn on the meter. 
f. 

Scan your operator barcode or enter your 5 digit Lawson ID. You must enter a leading zero if 
you  have  a  4  digit  Lawson  number.  If  your  ID  is  not  on  record  in  the  laboratory  you  will  be 
locked out of the instrument. 

NOTE:

 If the operator ID you enter is not accepted, attempt to 

re-enter it. If it is still rejected, contact your supervisor or Point of Care Coordinator.  

DO NOT

 

atte

mpt to perform tests under another operator’s ID. 

g.  From the 

Main Menu

, touch 

Patient Test

.  

h. 

Scan the patient’s financial number from the bar-coded wristband for the patient ID.  

i.  In any situation where a patient has not been registered, the number 0911911 should 

be used as the patient identification number.  

ii.  A custom comment should be entered in the ACCU-Check Inform ll system with the 

patient’s  name  at  the  time  of  testing.  This  will  allow  for  charting  and  billing  when  a 
financial number is assigned in the future. 

iii.  In  certain  situations,  an  employee  may  need  to  have  a  glucose  test  done.  If  this 

occurs,  the  number  0123456  should  be  used.  This  will  not  be  documented  as  a 
patient test. 

i. 

Confirm that the meter is coded (calibrated) to the same test strip code that is printed on the 
test  strip  vial  by  scanning  the  test  strip  vial. 

 

Contact  the  laboratory  if  you  are  unable  to 

confirm the correct test strip code. 

j. 

You  will  now  see  a  picture  of  a  test  strip  with  a  downward  flashing  arrow  on  the  screen 
indicating that you are ready to insert a test strip into the meter.  

k.  Remove a test strip from the vial and immediately recap the vial. Insert the test strip into the 

meter in the direction of the 

arrows and with the “ACCU-CHEK” lettering facing upward.  The 

Summary of Contents for Inform II Competency

Page 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Page 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Page 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Page 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Page 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Page 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Page 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Page 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Page 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Page 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Page 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Page 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Page 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Page 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Page 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Page 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Page 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Page 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Page 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Page 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Reviews: