Blood Glucose Monitoring System
New bio-sensor technology : 
Fast & Accurate results in 9 seconds

Blood Glucose Monitoring System for selftesting.

IMC00630  Rev.2014-05-03

Rm 1112, Ace Tower 9th Bldg., 130, Digital-ro,

Geumcheon-gu, Seoul, Korea

          

    

infopia Co.,Ltd.

132, Anyangcheondong-ro, Dongan-gu, 

Anyang-si, Gyeonggi-do, Korea (431-836)

http://www.infopia21.com

          

                      

Obelis S.A.

Bd. General Wahis 53 1030 Brussels, Belgium

Tel:+(32) 2.732.59.54 Fax:+(32) 2.732.60.03

E-mail:[email protected]

INFM16B

Lancet : infopia Co.,Ltd.

Lancing Device : GMMC

Parc Industrial Ciutat de Carlet Mestral 20, 46240

Carlet, Valencia, Spain

Tel : +34 963 691 484

Fax : +34 963 691 484

E-mail : [email protected]

Global Medical Market Corporation S.L.

Summary of Contents for EASYGLUCO Auto-coding

Page 1: ...nfopiaCo Ltd 132 Anyangcheondong ro Dongan gu Anyang si Gyeonggi do Korea 431 836 http www infopia21 com ObelisS A Bd GeneralWahis531030Brussels Belgium Tel 32 2 732 59 54Fax 32 2 732 60 03 E mail mai...

Page 2: ...Blood Glucose Monitoring System for selftesting...

Page 3: ...O Auto codingTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the three year warranty card and send it to us This will ensure tha...

Page 4: ...Auto codingTM Blood Glucose Monitoring System KIT Contents 8 EASYGLUCO Auto codingTM 9 r e t e M e s o c u l G d o o l B EASYGLUCO Auto codingTM Blood Glucose Meter Display 10 EASYGLUCO Auto codingTM...

Page 5: ...Month and Year 2 Test strip ejector Your Blood Glucose Monitoring System has been sealed and should contain all of the parts shown above Please return the system to the place of purchase or contact yo...

Page 6: ...YGLUCO Auto codingTM Test Strip vials in a cool dry place between 2 30 C Do not freeze 3 Make a notation of the discard date on the vial label when you first open it Discard remaining EASYGLUCO Auto c...

Page 7: ...10 000 feet will have an effect on test results Doctors Please note the following interferences that may affect test results test strip results can be used with hematocrit levels in the range of 20 t...

Page 8: ...ter re setting the test unit See also conversion table on page 38 2 Without setting the date properly the average glucose level and the results in memory will not show proper values It is recommended...

Page 9: ...he test strip vial If the control solution test results fall outside the range repeat the test Results that fall out side the range may be caused by Error in performing the test Failure to shake the c...

Page 10: ...ontrol Solution test results continuously falls outside the range printed on the vial the EASYGLUCO Auto codingTM system may not be functioning properly DO NOT use the system to test your blood until...

Page 11: ...cet in a biohazard garbage container Sterile lancets is used with a spring operated lancing device COMPONENTS Depth Indicator Adjustable Tip Lancet Holder Clear Tip Ejection System Sliding Barrel Rele...

Page 12: ...direction of arrow on test strip arrow up Please do not insert the test strip upside down When you insert the test strip into the meter the power automatically turns on with the code and temperature...

Page 13: ...d the meter turns off automatically Testing Procedure Caution 1 If the countdown does not start do not add more blood to the test strip Discard the test strip an restart testing 2 If you do not conduc...

Page 14: ...setup press and button to select the correct User ID then press caution 1 If you did not activate the User Activity option and or the User ID option Step 4 1 and or Step 4 2 procedures are not conduct...

Page 15: ...alues can be recalled in the LCD display at any time The newest result will replace the oldest beyond the 365 data stored Press shortly If the User ID option has been disabled during setup the newest...

Page 16: ...er significantly from fingertip samples The conditions in which these differences are more likely to occur are when your blood glucose is changing rapidly such as following a meal as insulin dose or a...

Page 17: ...d sample is taken If there is not enough blood gently massage the area until a sufficient sample has been collected caution To receive accurate test results at least 1 0 minimum sample volume is requi...

Page 18: ...o not refrigerate Avoid exposure to sun 4 Close the vial cap tightly immediately after removing a EASYGLUCO Auto codingTM strip Please refer to your EASYGLUCO Auto codingTM strip instruction if you wa...

Page 19: ...codingTM meter uses two 3V Lithium battery CR2032 The batteries are included in the set box When replacing the batteries only CR2032 or equivalent lithium batteries can be used Before replacing the b...

Page 20: ...ccuracy of your test strip by performing a control solution test After re test with blood sample two or three times When HI message persists please consult a doctor for assistance Problem with the tes...

Page 21: ...electrical current that will stimulate a chemical reaction This reaction is measured by the EASYGLUCO Auto codingTM Meter and displayed as your blood glucose result Note Different levels of reactions...

Page 22: ...20 0 250 260 270 288 300 306 325 342 360 20 8 22 2 23 0 24 0 25 0 26 4 27 7 30 0 33 3 375 400 414 432 450 475 500 540 600 Sample type Sample volume Test Range Reading time Calibration Operating Tempe...

Page 23: ...in the European community Three Year Warranty If at any time during the first three years after purchasem the meter does not work for any reason with the exception of obvious abuse misuse or disastrou...

Page 24: ...46 47 Memo Memo...

Page 25: ...48 Memo IMC00630 Rev 2014 05 03 INFM16B...