Symbols, error messages and troubleshooting
13
If this happens ...
it means:
When the meter is turned
on, the date is shown as
0- 0 and the time as 0:00
or 0:00
am
.
O
The meter was exposed to a temper-
ature below -10 °C and the battery is
beginning to freeze.
Eject the test strip or turn the meter
off. Move to a place where the tem-
perature is b10 °C and
+40 °C and wait for the meter to
warm up.
O
You have not set the time and date.
Set the time and date as described in
Chapter 3.2.4 “Setting the time-and-
date-format, time and date”.
Summary of Contents for GO
Page 1: ...BLOOD GLUCOSE METER Instructions for use ...
Page 42: ...42 Entering settings Month Day ...
Page 66: ...66 Entering settings Entering alarm times Hours Minutes Alarm 2 ...
Page 187: ...187 Patents 18 18 Patents US 5 463 467 US 5 366 609 US 6 707 554 ...