132
Farbe: P 249 – Prüfmittelnummer: 126 X
Symbols, display messages and troubleshooting
133
Symbols, display messages and troubleshooting
10
Symbol...
Meaning:
mem
signifies that the result being
displayed is a result from memory.
No results have been saved.
The meter cannot calculate the aver-
age of the last 7, 14 or 30 days,
as memory only contains results
that are not used in the calcula-
tion, e.g. quality control results,
as you have not entered the date
and time,
as you have set the date to a
time in the past.
The average blood glucose result for
the last 7, 14 or 30 days is displayed.
It was calculated using (for example)
24, 43 and 70 blood glucose results,
respectively.
Symbol...
Meaning:
You are requested to run a quality
control test.
This result is/was a quality control result.
The temperature during testing was at
the limit of the permitted range (between
+6 °C and +10 °C or b40 °C
and +44 °C).
Do not use the result as a basis for
therapeutic decisions.
Turn the meter off. Let the meter
warm up or cool down slowly. Turn
the meter on again and repeat the
test.
During testing the permitted tem-
perature range is from +10 °C to
+40 °C.