Symbols, display messages and troubleshooting
155
Symbols, display messages and troubleshooting
10
This message/symbol
appears…
and means:
You pulled the test strip out of the
meter before applying blood or con-
trol solution to the strip.
Press the Test button to turn the
meter off and then on again. Repeat
the test with a new test strip. Always
remove the test strip by pressing the
Test button to turn the meter off.
The test strip that was advanced is
defective.
Hold the meter upright so the test
strip is pointing downwards. Press
the Test button to turn the meter off
and allow the test strip to drop out.
Press the Test button to turn the
meter on again and repeat the test.
This message/symbol
appears…
and means:
There is condensation in the meter.
Press the Test button to turn the
meter off. Leave it to dry at room
temperature.
The batteries are empty or a meter
error has occurred.
Press the Test button to turn the me-
ter off and then on again. If the error
message appears again, the batter-
ies may be empty. Insert new bat-
teries. If you see the error message
again when you next turn the meter
on, please contact the Accu-Chek
Customer Careline (see Chapter 16
“Local customer support and ser-
vice”).