73
5
5.2 Troubleshooting Communication Between the Meter and
Pump
Display Shows
Possible Cause(s)
Possible Solution(s)
Bluetooth
icon fl ashing
j
Communication ended due to a
button press on the pump.
j
Meter and pump are outside
communication range.
j
Check the pump and continue operation
of the pump manually.
j
Ensure the pump has
Bluetooth
wireless
technology turned on.
j
Ensure the pump is within
communication range.
Bolus Advice
Bolus Advice data deleted
j
Bolus advice data integrity cannot
be confi rmed.
j
Wait 8 hours for accurate bolus advice.
IMPORTANT:
• Insulin doses and meals taken before the Bolus Advice warning are no longer
refl ected in the Bolus Advice calculation.
• The meter may not refl ect the pump bolus history, however the bolus history
is available on the pump.
45676_adv.indb 73