74
Display Shows
Possible Cause(s)
Possible Solution(s)
Bolus Delivery Unavailable
j
The meter cannot communicate
with the pump.
j
Retry or set to Manual Pump.
j
Ensure the meter and pump have
Bluetooth
wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on
the meter, the meter cannot
communicate with the pump. Replace
the batteries.
j
The pump is currently in Stop
mode.
j
Take the pump out of Stop mode.
j
The pump is currently delivering a
bolus.
j
Allow the current bolus delivery to
complete before delivering the next
bolus.
45676_adv.indb 74
11/3/09 4:45:22 PM
Summary of Contents for Aviva
Page 2: ...LAST UPDATE 2009 07 45676_adv indb 2 11 3 09 4 45 05 PM...
Page 6: ...4 45676_adv indb 4 11 3 09 4 45 05 PM...
Page 8: ...6 45676_adv indb 6 11 3 09 4 45 05 PM...
Page 26: ...24 45676_adv indb 24 11 3 09 4 45 10 PM...
Page 34: ...32 45676_adv indb 32 11 3 09 4 45 10 PM...