5. Gently squeeze the bottle to form one small drop.
6.
Bring the drop to the edge of the test strip within
the curve, and allow the test strip to automatically
draw the solution into the yellow window until it is
completely filled.
An hourglass symbol flashes
in the display until the
measurement is complete.
7. Tightly replace the cap on
control solution.
8. The control result appears. If result is within the
acceptable range (which is found on the side of
the container of test strips), a control bottle symbol
appears on the screen while the control test result
and time and date (if set) alternate with “OK” and
“L1” or “L2” on the display.
Example of an acceptable Level 1 control result.