Performing a Quality Control Check
1. Open the test strip container and remove a test strip.
Replace the container cap.
2. Within three minutes,
insert the test strip (yellow
window facing up) into test
strip slot. The meter turns
on automatically.
Check that the code
number on the meter
matches the code on the
container of test strips you
are using. The test strip
symbol stops flashing and
a blood drop symbol appears on the display.
3. Press and release the right arrow button once to
select a L1 (Level 1) control or twice to select a
L2 (Level 2) control.
Note: Results marked as L1 or L2 do not appear
in memory or averages.
4. Hold the control bottle
horizontally with the tip
pointed directly at the right
edge of the test strip.
24