28 OMNISOUND
®
3000C PRO USER MANUAL
COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Effective Radiating Area (ERA) and Beam Non-Uniformity Ratio (BNR)
SPATIAL PATTERN
The applicator produces a collimated cylindrical beam when measured at 5mm from the radiating surface with
distilled degassed water 30
o
C with line voltage variations of + 10 percent of rated value for the 5cm
2
transducer at 1
and 3MHz and the 2cm
2
transducer at 3MHz. At 1MHz the 2cm
2
transducer produces a diverging beam. The graph
on the following page is representative of a three dimensional plot of the beam at 5mm from the transducer surface.
POWER AND INTENSITY
Output Power
2cm
2
5 cm
2
1MHz
0-3W
0-12W
Nominal
3MHz
0-3W
0-9W
Nominal
Output Intensity
1MHz
0-2W/cm
2
0-3W/cm
2
Nominal
3MHz
0-2W/cm
2
0-2W/cm
2
Nominal
CERTIFICATION
The Omnisound
®
complies with the ultrasound performance standards as set forth in 21 CFR 1050.10.
Patented in USA Nos.: 5,086,788; 5,413,550; D 315,958; other patents pending.
ABBREVIATIONS
RTPA
MAX/AVG INTENSITY RATIO,
MHz
Megahertz
Ratio of Temporal-Maximum Effective
cm
2
Square Centimeter
Intensity to the Temporal-Average
CW
Continuous Wave
Effective Intensity
W
Watts
Gen
Generator
W/cm
2
Watts per Square Centimeter
S/N
Serial Number
P.P.S.
Pulses per Second
Freq.
Frequency
ERA
Effective Radiating Area
BNR
Beam Non-Uniformity Ratio
Coll.
Collimating