Effective Radiating Area (ERA) and Beam Non-Uniformity Ratio (BNR) 

 

SPATIAL PATTERN 

The  applicator  produces  a  collimated  cylindrical  beam  when  measured  at  5mm  from  the  radiating  surface  with 
distilled degassed water 30

o

C with line voltage variations of + 10 percent of rated value for the 5cm

transducer at 1 

and 3MHz and the 2cm

2

 transducer at 3MHz. At 1MHz the 2cm

2

 transducer produces a diverging beam. The graph 

on the following page is representative of a three dimensional plot of the beam at 5mm from the transducer surface. 

  

POWER AND INTENSITY

 

Output Power

 

2cm

2

 

5 cm

2

 

 

1MHz 

0-3W 

0-12W 

Nominal 

3MHz 

0-3W 

0-9W 

Nominal 

Output Intensity 

 

 

 

1MHz 

0-2W/cm

2

 

0-3W/cm

2

 

Nominal 

3MHz 

0-2W/cm

2

 

0-2W/cm

2

 

Nominal 

 

CERTIFICATION

 

The Omnisound

®

 complies with the ultrasound performance standards as set forth in 21 CFR 1050.10.  

Patented in USA Nos.: 5,086,788; 5,413,550; D 315,958; other patents pending.

 

 

ABBREVIATIONS

 

RTPA

  MAX/AVG INTENSITY RATIO,   

 

MHz

 

Megahertz 

 

Ratio of Temporal-Maximum Effective  

 

cm

2

 

Square Centimeter 

Intensity to the Temporal-Average   

 

CW

 

Continuous Wave 

Effective Intensity 

 

 

 

W

 

Watts 

Gen

 

Generator 

 

 

 

 

W/cm

2

  Watts per Square Centimeter 

S/N

 

Serial Number 

 

 

 

 

P.P.S.

  Pulses per Second 

Freq.

 

Frequency 

 

 

 

 

ERA

 

Effective Radiating Area 

BNR

 

Beam Non-Uniformity Ratio 

 

 

Coll.

 

Collimating 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for 1203000C

Page 1: ......

Page 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Page 3: ......

Page 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Page 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Page 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Page 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Page 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Page 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Page 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Page 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Page 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Page 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Page 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Page 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Page 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Page 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Page 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Page 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Page 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Page 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Page 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Page 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Page 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Page 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Page 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Page 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Page 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Page 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Page 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Page 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Page 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Page 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Page 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Page 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Page 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...

Page 37: ......

Reviews: