OMNISOUND
®
3000C PRO USER MANUAL
11
COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
When the transducer output is switched from Main to Aux output with the transducer select switch, the auto retest
screen will start up to ensure proper function of the Omnisound
®
3000C Pro system to the new transducer.
Additional Tests/Warnings
If the sensor in the transducer head detects a temperature over 47
o
C (116.6
o
F), the word "Over Heat" will flash on
the display and the unit will emit three "beeps." The unit will become inoperable and the treatment time will be
temporarily frozen until such time as the temperature drops to the proper level.
When the sensor checks the temperature and finds that it has returned to the proper level, the unit will emit two
"beeps," and the word "Over Heat" will disappear and the display screen will return to normal. The user may then
press START to continue the treatment with the existing parameters. Alternatively, the treatment time and intensity
may be reset before pressing START to recommence the treatment.
NOTE:
Make sure that the output is switched OFF when the transducer is not in use. Temperature will rise
rapidly if the output is switched ON and the transducer is not in contact with the skin.
Power Supply Operation
Omnisound
®
3000C Pro is a line power only device. It is not intended for use with batteries.
Operational Sequence
Connection of the Transducer
Select the desired transducers (2cm
2
or 5cm
2
) for use with the Omnisound
®
3000C Pro. Position the transducer
cable plugs so that the 12 prongs align correctly with the 12 receptor holes in the transducer cable MAIN socket or
AUX socket. This can be accomplished quickly by placing the inside key at the 12 o’clock position. Insert the
transducer cable plugs into the transducer cable sockets and turn the knurled cuff of the plug clockwise many times
until the threaded barrel of the plug is no longer visible. The transducers should now be firmly attached to the
Omnisound
®
3000C Pro
generator.
Connecting Line Power
Insert the three-prong hospital grade plug of the AC line cord into a wall outlet supplying a 100 - 240 volt,
alternating current, 47-63 Hertz frequency. Plug in the other end (female connector) into the supplied AC power
supply. Insert the power supply output connector into the receptacle on the Omnisound
®
3000C Pro
generator
located next to the power switch on the back of the unit. After the transducers are connected the power may be
applied by pressing the power switch to the ON position.