Accelerated Care Plus 1203000C User Manual Download Page 14

 OMNISOUND

®

 3000C PRO USER MANUAL 

 

 11 

 COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED

 

 
When the transducer output is switched from Main to Aux output with the transducer select switch, the auto retest 
screen will start up to ensure proper function of the Omnisound

®

 3000C Pro system to the new transducer. 

 

Additional Tests/Warnings 

 
If the sensor in the transducer head detects a temperature over 47

o

C (116.6

o

F), the word "Over Heat" will flash on 

the  display  and  the  unit  will  emit  three  "beeps."  The  unit  will  become  inoperable  and  the  treatment  time  will  be 
temporarily frozen until such time as the temperature drops to the proper level. 
 
When  the  sensor  checks  the  temperature  and  finds  that  it  has  returned  to  the  proper  level,  the  unit  will  emit  two 
"beeps," and the word "Over Heat" will disappear and the display screen will return to normal. The user may then 
press START to continue the treatment with the existing parameters. Alternatively, the treatment time and intensity 
may be reset before pressing START to recommence the treatment. 

 

NOTE:

  Make sure that the output is switched OFF when the transducer is not in use. Temperature will rise 

rapidly if the output is switched ON and the transducer is not in contact with the skin. 

 

Power Supply Operation 

 

Omnisound

®

 3000C Pro is a line power only device.  It is not intended for use with batteries. 

 

Operational Sequence 

 

Connection of the Transducer 

 
Select  the  desired  transducers  (2cm

2

  or  5cm

2

)  for  use  with  the  Omnisound

®

  3000C  Pro.    Position  the  transducer 

cable plugs so that the 12 prongs align correctly with the 12 receptor holes in the transducer cable MAIN socket  or 
AUX  socket.  This  can  be  accomplished  quickly  by  placing  the  inside  key  at  the  12  o’clock  position.    Insert  the 
transducer cable plugs into the transducer cable sockets and turn the knurled cuff of the plug clockwise many times 
until  the  threaded  barrel  of  the  plug  is  no  longer  visible.  The  transducers  should  now  be  firmly  attached  to  the 
Omnisound

®

 3000C Pro

 

generator. 

 

Connecting Line Power 

 
Insert  the  three-prong  hospital  grade  plug  of  the  AC  line  cord  into  a  wall  outlet  supplying  a  100  -  240  volt, 
alternating  current,  47-63  Hertz  frequency.  Plug  in  the  other  end  (female  connector)  into  the  supplied  AC  power 
supply.  Insert  the  power  supply  output  connector  into  the  receptacle  on  the  Omnisound

®

  3000C  Pro

 

generator 

located  next  to  the  power  switch  on  the  back  of  the  unit.  After  the  transducers  are  connected  the  power  may  be 
applied by pressing the power switch to the ON position. 

Summary of Contents for 1203000C

Page 1: ......

Page 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Page 3: ......

Page 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Page 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Page 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Page 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Page 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Page 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Page 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Page 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Page 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Page 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Page 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Page 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Page 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Page 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Page 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Page 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Page 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Page 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Page 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Page 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Page 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Page 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Page 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Page 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Page 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Page 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Page 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Page 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Page 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Page 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Page 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Page 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Page 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...

Page 37: ......

Reviews: