Accelerated Care Plus 1203000C User Manual Download Page 12

 OMNISOUND

®

 3000C PRO USER MANUAL 

 

 9 

 COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED

 

 

M.)  Transducer Data 

 

E.)  Frequency 

 

D.)  Duty Factor 

H.

 

The CONTRAST button is used to lighten or darken the digital screen contrast. 

I.

 

The transducer SELECT button will alternate selected transducer from MAIN or AUX. 

J.

 

Transducer cable plug-in socket used to connect the transducer to the unit.  The socket marked as MAIN must 
be used when only one transducer is used with the unit. 

K.

 

Transducer cable connector inserts into the plug-in socket for connection of the transducer to the unit. 

L.

 

Transducer coaxial cable. 

M.

 

Pressing  this  button  displays  the  transducer  BNR  and  ERA.    Pushing  ASSIST  button  followed  by  this  button 
will display the date of last calibration. 

N.

 

The OUPUT button is used to enable power calibration of the generator and the selected transducer. 

O.

 

This button is used to set the TEMP mode, and activate the Delta T mode. 

P.

 

The POWER/INTENSITY button is used to toggle the display in watts (power) or watts/cm

² 

(intensity). 

Q.

 

The cooling fan provides airflow through the unit when the unit is operational. 

R.

 

The ELECTROSTIM INPUT PORT is used to connect the shrouded pin-to-banana adaptor (

Item No. 2503036

and lead wire from electro-stimulator for combination therapy. 

S.

 

Toggle POWER switch;  turn on AC power when set to 1, and turns off the AC power when set to 0. 

T.

 

The carry handle for easy transport of the equipment. 

U.

 

AC line cord receptacle, where the power cord is plugged in. 

V.

 

The generator CAUTION LABEL is located on bottom of the unit. 

W.

 

Two TRANSDUCER HEADS come standard with the unit (both 2 and 5 cm

²

), and both are adjustable to any 

angle from 0 to 360

º

X.

 

The TRANSDUCER LABEL lists the transducer serial number, BNR and ERA data. 

Y.

 

The FREQUENCY button is used to select: 

 

1 MHz for deep tissue penetration, or 

 

3 MHz for superficial tissue penetration. 

Z.

 

The DELTA T temperature mode select button is used to activate temperature calculation mode. 

AA.

 

The POWER intensity set button is used to increase and decrease the ultrasound output level. 

BB.

 

This  LED  is  on  continuously  to  show  which  transducer  is  selected.    The  LED  flashes  on  and  off  when 
overheated. 

 

 

Summary of Contents for 1203000C

Page 1: ......

Page 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Page 3: ......

Page 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Page 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Page 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Page 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Page 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Page 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Page 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Page 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Page 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Page 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Page 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Page 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Page 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Page 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Page 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Page 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Page 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Page 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Page 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Page 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Page 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Page 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Page 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Page 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Page 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Page 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Page 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Page 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Page 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Page 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Page 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Page 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Page 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...

Page 37: ......

Reviews: