Accelerated Care Plus 1203000C User Manual Download Page 27

24                                                                                                                                      OMNISOUND

®

 3000C PRO USER MANUAL 

 

 

 

COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

Temperature Mode of Operation / Delta T Mode 

 
In  order  to  achieve  repeatable  thermal  effects  of  ultrasound  in  tissue  the  dosage  of  the  ultrasound  beam  must  be 
carefully controlled. This involves accurate determination  of the treatment area, couplant to be used,  frequency of 
operation, BNR and the desired temperature increase required to achieve appropriate therapeutic effects in the tissue. 
 
LEHMAN and other investigators have categorized the effects of ultrasound on tissue as being of a mild or 1 degree 
C  increase,  moderate  or  2  degree  C  increase  or  vigorous  or  4  degree  C  heating  effect  above  the  baseline  tissue 
temperature  of  37.5  degrees  C.  Typically  vigorous  heating  is  used  to  treat  connective  tissue  contracture  and  scar 
tissue while moderate and mild heating are used in pain control, sub acute and chronic inflammatory conditions and 
for muscle spasm reduction. 
 
The  patented  Delta  T  mode  of  the  Omnisound

®

  3000C  Pro  is  a  proprietary  software  system  that  is  used  in 

conjunction with the ultrasound system to ensure consistent dosimetry. 
 
To use the system proceed as follows: 
 
1.

 

After the unit has completed self-test and the default screen appears depress the TEMPERATURE MODE SET 
UP button on the generator or the ΔT button on the transducer.  The screen will change to the temperature mode 
select screen.  Then press the SELECT button to select the Delta T mode desired. 

2.

 

Select the couplant being used, if other than ultrasound gel, by using the set parameter switches (normally this is 
not used when ultrasound gel is selected). 

3.

 

Select the frequency of 1 or 3MHz.

 

NOTE:

  3MHz is the default frequency so if the depth is less than 1" or 2.5cm for the target tissue then you will 

not need to change the frequency. If the depth is greater than 1", select the 1MHz frequency. 

4.

 

Press the ΔT button on the transducer, or the TEMP MODE SETUP button on the screen to go to the main ΔT 
run screen.  Default is 4 ΔT. 

5.

 

To start the treatment, supply adequate couplant, apply the transducer to the patient and increase the power. As 
you do so you will notice that the timer will change from 15 minutes to a lower value. This will only occur if 
the  dose  is  sufficient  to  reach  the  desired  temperature  within  the  treatment  time.  Adjust  the  output  intensity  - 
typically  to  a  range  of  0.5W/cm

2

  to  1W/cm

2

 

is  used  based  on  patient  comfort.  The  treatment  time  will 

automatically adjust based on the power setting chosen to provide the correct dosage. As the power is increased 
the  time  decreases  and  vice  versa.  If  the  patient  experiences  any  discomfort  decrease  the  power  by  a  small 
amount  (0.1  -  .2W/cm

2

)  and  note  if  the  pain  is  reduced.  If  so  continue  the  treatment,  otherwise  decrease  the 

power  again.  The  treatment  time  will  automatically  increase  to  compensate  for  the  decreased  power  level. 
Repeat again if the sensation is still too hot following the reduction of power. 

NOTE:

    The  delta  T  mode  of  operation  is  calibrated  for  muscle  not  in  proximity  of  bone.  If  the  ultrasound 

application at 3MHz is within 1 cm of bone the 2 degree delta T mode will heat at a rate of 4 degrees, or the 1 
degree delta T mode will heat at 2 degrees. Within .5cm proximal to bone, the 1 degree Delta T mode will heat 
at 4 degrees due to the reflection of the ultrasound field from the bone back into the overlying tissue. 

6.

 

The temperature display shows the projected temperature  increase over time based on the selected parameters 
for  treatment.  Although  it  does  not  represent  actual  temperature  it  should  be  a  close  approximation  if  the 
treatment is correctly applied. 

7.

 

Use  a  slow  (approximately  1cm  per  second)  linear  or circular  motion  with  the  transducer. The  treatment  area 
should be restricted to two of the transducer sizes, which approximates 2 effective radiating areas (ERA). This 
is absolutely mandatory if the correct relationships in the calculation equations are to be met. 

8.

 

When the treatment is complete the timer will beep, the output will power down. To treat another area of 2 ERA 
simply push the up power control on the transducer and the mode will be set up automatically at the originally 
set default parameters. To change the defaults select a new protocol from the front panel switches. 

Summary of Contents for 1203000C

Page 1: ......

Page 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Page 3: ......

Page 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Page 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Page 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Page 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Page 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Page 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Page 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Page 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Page 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Page 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Page 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Page 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Page 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Page 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Page 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Page 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Page 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Page 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Page 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Page 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Page 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Page 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Page 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Page 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Page 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Page 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Page 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Page 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Page 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Page 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Page 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Page 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Page 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...

Page 37: ......

Reviews: