Handheld pulse oximeter user’s manual
Guidance and Declaration – Electromagnetic Immunity
The device is suitable for use in the electromagnetic environment specified below. The
customer or the user of the device should make certain that they are in such an
environment when operating it.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV
contact
±
8 kV air
±
6 kV
contact
±
8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient / burst
(EFT)
IEC 61000-4-4
N/A
N/A
N/A
Surge IEC
61000-4-5
N/A
N/A
N/A
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
N/A
N/A
N/A
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic The
power frequency of magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Declaration – electromagnetic Immunity
11