3.15 Setting the channel range
Your Control Medi has the option to set different channel ranges for the transmitter to operate within.
When the change of channel range is applied, programmed channels will disappear and a new
random code is added on the
Code
button
!
.
Normally, the channel range does not need to be changed. If several GewaLink transmitters are to
be used in the same area and you do NOT want to control the same equipment, it may be
necessary to use different channel ranges.
Table 3 shows the different settings.
Table 3: Channel range
Channel range, GewaLink channels
00-23
1
16-39
2
32-55
3
56-79
4
80-103
5
104-127
6
Level 1
00-05
16-21
32-37
56-61
80-85
104-109
Level 2
06-11
22-27
38-43
62
1
-67
86-91
110-115
Level 3
12-17
28-33
44-49
68-73
92-97
116-121
Level 4
18-23
34-39
50-55
74-79
98-103 122-127
1
Channel 62 is a stop code and cannot control any function. On the other hand you can store channels/codes
on this button which in turn can control a function.
Do as follows to set the channel range:
1. First press
---
and then simultaneously press
2
for 2 seconds. When you release the buttons the
“Setting” alert sounds and the indication lamp on button 2 flashes.
2. Press
4
.
3. Choose the channel range
1
to
6
according to table 3 above.
4. If the action is successful, the “Success” alert sounds. If something goes wrong or takes too long,
the “Error” alert sounds.
Setting the channel range
Press
---
+
2
2 sec
,
4
,
choose the channel range
.
1
=
00-23,
2
= 16-39,
3
= 32-55,
4
= 56-79,
5
= 80-103,
6
= 104-127
Default setting: GewaLink channel 16-39
Summary of Contents for Control Medi Standard
Page 1: ...Control Medi Standard Manual Art Nr 429220 1 SE EN 4 0...
Page 2: ......
Page 5: ...11 Summary and Quick guide 49...
Page 10: ......
Page 28: ......