Position 2
(guide X0 Y0 Z340)
Correction calculation for XY compensation
(Pos 1 + Pos 2)/2
=
Average length and our target length
(Pos 1 / average)-1 =
compensation (could be + or -, try + first if error doubled
reverse the polarity.
*If you click on test map and apply, the measurement will update on the screen and show
if correction is the correct polarity. If wrong, reverse sign and click on test map button and
apply again. Once happy, re-measure spheres.
Position 3
XZ
(guide X0 Y0 Z285)
When measuring the lower sphere, take at least 7 points, measuring as much surface as
possible, this will produce a better centre co-ordinate.
Due to the other spheres still
within the screen inspection, the
safe volume may prevent the
program from running. Instead
use the play button within the
sphere measure window to
apply controlled cnc moves.
Make sure probe can move to
first point safely.