X axis
Position 1
Position the tip as shown on the corner of the slip, (guide X110 Y85, Z325)
Measure plane to plane, or plane to point. Dimension this on the screen, enter the
nominal gauge size, with a tolerance of 0.0038mm, this represents 0.003mm base
tolerance plus 0.4 microns for each 100mm of gauge length, hence 200mm would be
additional 0.8 microns).
At this stage we do not compensate for linear error
Position 2
(guide X110 Y-100 Z325)
In the event
that yaw Xdy compensation is required…
Correction = (pos 1 result
– pos 2 result) / 200
Divide this result by distance between pos 1 and pos 2. Approx. 185