Once homed the panel should look like this
Once homed check the Z counterbalance. Click on Assess button and click ok to prompt.
Make sure joystick speed paddle is not on zero. Z will move up and down, no need to edit
Z Min or Z Max. On completion a number is displayed. Target is 0.000 +/- 2.00, to adjust
loosen or tighten the screw attached to the counterbalance spring, and click Assess again
to see affect of adjustment.
Current strut temperature,
target is 40 degrees
Strut state, if error code
place mouse pointer over
number to see error
Heat status
– Heating is
stopped once 40 degrees is
achieved.
Encoder signal strength
PWM - Power consumption, from version 32, a
lazy mode is active (after 30 secs). If machine
is idle the machine will move a very small
amount to relieve the power consumption of
the strut, as its position governs the power
used. This action prevents struts overheating.