Temp sensor settings as below (do not choose a machine)
By clicking on sensor icon or by clicking on the temperature display at the base of the
screen, will open the screen below. Tick
‘Use Temperature Compensation’, also by
clicking on the down arrow for ‘Extended Temperatures’ you can see the individual strut
temperatures, target is 40 deg. C. If not metrology ready is flashing on the screen, it
means temperature is not yet reached, moving the machine will quickly attain temperature.
Step 15
It is possible to fine tune the
ambient temperature sensor, by
entering an offset here. Use a
separate temperature gauge to
determine actual room
temperature.