Menu / parameter
Description
Diagnostics / . . .Diagnosis Control / . . .Noise Check
Start Noise Check
Start the ‘
Noise Check
’ function, using .
Power Spectrum
Current power spectrum.
Amplitude 1 Value
Display the four highest amplitudes in the power spectrum.
Amplitude 2 Value
Amplitude 3 Value
Amplitude 4 Value
Frequency 1
Display the corresponding frequency to those four highest amplitudes in the power spectrum.
Frequency 2
Frequency 3
Frequency 4
Diagnostics / . . .Diagnosis Control / . . .Fingerprints
Tx Factory CMR
The "fingerprint" database allows to compare the values at the time of factory calibration with the currently
recorded values. Errors in the integrity of the "Tx Factory 1m/s" device can be detected at an early stage.
Corrective actions can be taken.
Tx Factory 1m/s
Tx Factory 10m/s
Start Tx Customer FP
Create the manual fingerprint for the transmitter using .
Tx Customer CMR
The manual fingerprint is created on site prior to verification of the transmitter.
Use the parameter "Start Tx Customer FP" to create the the manual fingerprint.
Tx Customer 1m/s
Tx Customer 10m/s
7 . . .Operation
. . .Parameter descriptions
. . .Menu: Diagnostics