1
2
1
2
3
Transmitter
FET612
Remote mount design
FEM612
Integral mount design
FEM611
5
FEM610 |
PROCESSMASTER | OI/FEM610-EN REV. A
2 Design and function
Overview
ProcessMaster
1
Flowmeter sensor 1/10" ...3/8" (DN3 ... DN10)
2
Flowmeter sensor 1/2"...4" (DN15...DN100)
3
Remote transmitter
Figure 1 Designs
Flowmeter sensor
Model
ProcessMaster FEM611, FEM612
Housing
Integral mount design, remote mount design
Measuring accuracy for liquids
0.5 % of measured value
Permissible measuring medium
temperature T
medium
−13 to 266 °F (−25 to 130 °C)
Minimum conductivity
> 20 µS/cm
Nominal pressure rating
ASME CL 150, CL 300
Nominal diameter
1/10“ ... 4“ (DN3 ... DN100)
Process connection
Wafer style connection
Liner material
ETFE
Electrode material
Hastelloy C®, Platinum-Iridium, Tantalum
IP rating
IP 65 / IP 67 / Type 4X for remote sensor assembly
Approvals for sensor
CRN (Canadian Reg.Number)
Pending
Further approvals
At www.abb.com/flow or on request.
Transmitter
Model
FET612
Housing
Integral mount design (see Fig. 1), remote mount design.
IP rating
IP 65 / IP 67 / Type 4X
Cable length
Maximum 164 ft (50 m), remote mount design only
Power supply
100 ... 240 V AC (-15 / +10 %),
24 ... 48 V DC (-10% / +10 %)
Outputs
Current output: 4 … 20 mA, active
Digital output 1: passive, configurable as pulse, frequency or switch output
Digital output 2: passive, configurable as pulse or switch output
Local display
Configurable graphical display (option)
Approvals for transmitter
Further approvals
At www.abb.com/flow or on request.