116
FEP630, FEH630
ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
… 9 Diagnosis / error messages
… Overview
Error no. / Range Error text
Current output
Digital output
Pulse output
LCD display
Error maskable?
S065.044 /
operation
Inhouse temp. alarm.
Reduce ambient temperature.
Current value - no
change.
No Answer
No Answer
Current value - no
change.
S064.041 /
operation
EPD alarm.
Secure pipe is completely filled.
Alarm - configured
as in menu ‘Curr. at
EPD Alarm’.
Alarm, if DO as
‘Logic / Alarm Signal
/ EPD Alarm’ is
configured
0 Hz
0 %
Menu ‘...Group
Masking’.
S063.040 /
operation
TFE alarm.
Secure pipe is completely filled.
Alarm - configured
as in menu ‘Curr. at
TFE Alarm’.
Alarm, if DO as
‘Logic / Alarm Signal
/ TFE Alarm’ is
configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S062.039 /
operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
Current value - no
change.
No Answer
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S061.038 /
operation
Conductivity limits alarm.
Change limits or Check
application.
Current value - no
change.
Alarm, if DO as
‘Logic / Alarm Signal
/ Conductivity’ is
configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S060.037 /
operation
Gas bubble alarm.
Check conditions of application.
Current value - no
change.
Alarm, if DO as
‘Logic / Alarm Signal
/ Gas bubble Alarm’
is configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S052.016 /
operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
Alarm - configured
as in menu
‘„Curr.Out > 20.5mA’.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S051.017 /
operation
Curr.Out V1 / V2, V3 / V4
saturated.
CO process value out of range.
Adapt Qmax.
S049.019 /
electronics
Option Card 1 com error.
Defective Card. Check Card 1.
Call Service.
Does not react
anymore
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S048.020 /
electronics
Option Card 2 com error.
Defective Card. Check Card 2.
Call Service.
Does not react
anymore
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.